Sensitivities of inductively coupled plasma optical emission spectrometry for dry and wet aerosols

Alfred P. Weber, R. Keil, Leo Tobler, Urs Baltensperger

Analytical Chemistry · 1992 · 24 citations · 48 references

Concepts

Abstract

The sensitivity of the Inductively coupled plasma optical emission spectrometry (ICP-OES) signal was Investigated for dry aerosols generated by spark discharges, as well as aerosols generated by a conventional ultrasonic nebulizer (USN). Particle sizes of the generated aerosols were &lt;300 nm and the ICP sensitivity was not Influenced by particle size. The small amount of residual water after the desolvator (1.5 mg/min) did not change the plasma conditions, since the excitation temperature In the plasma remained constant within 61 K for the dry and wet aerosols. Nevertheless, mixing of the dry aerosol with H<sub>2</sub>o/HNO<sub>3</sub> droplets resulted In a reduction of the ICP signal by 43.1% which could be attributed to scavenging of analyte particles by the droplets generated In the USN. An absolute calibration of the ICP-OES signal by neutron activation analysis was performed using silver aerosols. With these calibration data the nebulizing efficiency of the USN was determined, and a value of (5.2 ± 0.5)% was found.