Radiology · 2003 · 33 citations · 16 references
MedicineCardiologyRadiologyDiagnostic ImagingThoracic ImagingThoracic RadiologyAortic ImagingCardiac ImagingCardiovascular ImagingComputed TomographyComputed Tomography PhysicsElectrocardiographyAortic Motion AssessmentCardiac Computed TomographyCardiac Imaging ArtifactsComputed Tomography ArtefactsHeart RateMotion Artifact CorrectionMotion Artifact DetectionMotion Artifact Removal
The authors assessed motion artifact of the thoracic aorta in 25 patients who underwent multi-detector row computed tomography (CT) with retrospective electrocardiographic (ECG) gating. CT reconstructions centered at four phases of diastole were compared for five different levels of the thoracic aorta. A significant positive correlation was observed between heart rate and motion artifact (r = 0.72, P <.001). The optimal reconstruction phase varied between patients, and this was directly related to heart rate. For patients with a heart rate of 70 beats per minute, the reconstruction phase centered at 75% of the R-R interval had the significantly least motion artifact (P =.004). Conversely, the optimal reconstruction phase for patients with heart rates above 70 beats per minute was centered at 50% of the R-R interval (P =.09).
16
Coronary angiography with multi-slice computed tomography
Koen Nieman, Matthijs Oudkerk, Benno J. Rensing et al. · The Lancet · 2001 · 726 citations
Medicine, Cardiology, Radiology +12
Noninvasive Coronary Angiography by Retrospectively ECG-Gated Multislice Spiral CT
Stephan Achenbach, Stefan Ulzheimer, Ulrich Baum et al. · Circulation · 2000 · 435 citations · Full text
Medicine, Cardiology, Radiology +14