MedicinePediatricsNeonatologyPulmonologyVentilationBreathing SupportMechanical VentilationNeonatal VentilationPressure Support VentilationPressure-control VentilationCardiac OutputCardiac Output AugmentationCardiac Output VariabilityHeart RateMean Airway PressureMechanical Ventilation In NewbornsNeonatal HemodynamicsNeonatal Mechanical Ventilation
Fifteen neonates and infants were investigated during pressure controlled ventilation (PCV) and pressure support ventilation (PSV) on a Servo 300 ventilator. Changes in cardiac output (aortic mean blood flow velocity) were assessed with the Doppler technique. During PSV cardiac output increased by 16% (P < 0.01) compared with PCV with equal ventilation. Mean airway pressure decreased significantly during PSV compared with PCV, which may explain the increased cardiac output. During PSV, when inspiratory/expiratory (I:E) relations are dependent on the patient, we found a lower I:E ratio compared with PCV. It can be assumed that at least part of the decreased mean airway pressure is caused by the lower I:E ratio. As the heart rate was unchanged, the variations in cardiac output were caused by alterations in stroke volume. It is concluded that a patient triggered mode should be preferred providing that the neonate/infant has the ability to elicit the ventilator's triggering system.