Physics in Medicine and Biology · 2007 · 136 citations · 14 references
MedicineRadiation OncologyRadiologyRadiation DosimetryRadiation MeasurementTreatment PlanningDose CalculationDosimetry InstrumentationIonization Chamber DosimetryIonization ChambersQuality Assurance In DosimetryRadiation Dose MeasurementEnergy DependenceIonization Chamber DetectorsIonizing Radiation DosimetryLateral Dose ProfilesPatient Dose MeasurementPhoton Energy DependenceTreatment Planning System Dose VerificationTwo-dimensional Dosimetry
In this paper we present the results of a dosimetric evaluation of a 2D ionization chamber array with the objective of its implementation for quality assurance in clinical routine. The pixel ionization chamber MatriXX (Scanditronix Wellhofer, Germany) consists of 32x32 chambers with a distance of 7.6 mm between chamber centres. The effective depth of measurement under the surface of the detector was determined. The dose and energy dependence, the behaviour of the device during its initial phase and its time stability as well as the lateral response of a single chamber of the detector in cross-plane and diagonal directions were analysed. It could be shown, that the detector's response is linear with dose and energy independent. Taking the lateral response into account, two different dose profiles, for a pyramidal and an IMRT dose distribution, were applied to compare the data generated by a treatment planning system with measurements. From these investigations it can be concluded that the detector is a suitable device for quality assurance and 2D dose verifications.
14
Dosimetric IMRT verification with a flat‐panel EPID
Brad Warkentin, S Steciw, S Rathee et al. · Medical Physics · 2003 · 205 citations
Two‐dimensional ionization chamber arrays for IMRT plan verification
B Poppe, Arne Blechschmidt, Armand Djouguela et al. · Medical Physics · 2006 · 202 citations · Full text
Verification of IMRT fields by film dosimetry
M. Bucciolini, F. Banci Buonamici, M. Casati · Medical Physics · 2003 · 102 citations