Liquid Chromatographic Determination of Polar Pollutants in Surface Water Using Membrane Extraction Disks for On-Line Trace Enrichment

E. Brouwer, Dick J. van Iperen, Igor Liška, H. Lingeman, U.A.Th. Brinkman

International Journal of Environmental & Analytical Chemistry · 1992 · 50 citations · 5 references

Concepts

Abstract

Abstract Membrane extraction disks loaded with alkyl-bonded silica, copolymer or ion-exchange material have been tested for the on-line trace enrichment of polar pesticides from surface water. To this end, a new membrane extraction disk holder has been designed, which can be used in the forward flush and backflush mode. The basic compounds carbendazim, chloridazon, simazine and 4-chloroaniline were used as test compounds. They are preconcentrated from acidic solution using one or two membrane extraction disk holders in series (operated in the reversed-phase and ion-pair mode), and analysed by gradient liquid chromatography with polymer PLRP-S as the stationary phase and aqueous acetonitrile mixtures (pH 3) as the eluent. Data on the efficiency and repeatability of the on-line procedure are reported. Preconcentration of 30 ml of surface (river Rhine) water allows the UV detection of the four pollutants at the 0.1 μg 1−1 level.

References