Cardiac troponin T detection using polymers coated quartz crystal microbalance as a cost-effective immunosensor

Krongkamol Wong-ek, Orawon Chailapakul, Noppadon Nuntawong, Kata Jaruwongrungsee, Adisorn Tuantranont

Biomedizinische Technik/Biomedical Engineering · 2010 · 17 citations · 29 references

Concepts

Abstract

Cardiac troponin T (cTnT) detection has been the focus of increased interest due to its role in myocardial infarction diagnosis. In this study, we report a relatively low coat technique to detect cTnT using a quartz crystal microbalance (QCM) sensor. A sensitive detection is achieved by introducing a QCM surface with a carboxylic polyvinyl chloride immobilization layer. The surface morphologies of this polymer film under varied deposition thickness have been investigated by field emission scanning electron microscopy and atomic force microscopy. A cTnT detection result from a modified QCM surface can be obtained within a short response time by a direct detection of the immunoreaction and a direct conversion of mass accumulation into a frequency shift, representing a measurable electrical signal. The relationship between the cTnT concentration and the response current from a QCM sensor shows detectability at the concentration of cTnT as low as 5 ng/ml.

References

29