Ultrasonic computed tomography of the breast. Improvement of image quality by use of cross-correlation time-of-flight and phase-insensitive attenuation measurements.

Thomas L. Chenevert, Donita Irene Bylski, Paul L. Carson, Charles R. Meyer, Peyton H. Bland, Dorit D. Adler, R.M. Schmitt

Radiology · 1984 · 29 citations · 0 references

Concepts

Abstract

The authors present two methods of reducing refraction artifacts in ultrasonic computed tomography of the breast. Measuring the time of flight of sonic pulses by cross-correlation instead of leading-edge detection significantly reduces distortion and improves resolution in speed-of-sound images. Phase-insensitive reception of pulse energy across a large aperture array is shown to be superior to a conventional single-element transducer for attenuation imaging.