Physics in Medicine and Biology · 1983 · 55 citations · 9 references
MedicineRadiation OncologyRadiation DosimetryRadiation PhysicsRadiation TransportAbsorbed DoseBrachytherapy DosimetryBrachytherapy PhysicsMonte Carlo DosimetryPhoton DosimetryRadiation Dose MeasurementBrachytherapy Dose CalculationBrachytherapy Photon DosimetryDegradationDose Distribution CalculationsPhotoelectric EffectPhoton Dose DistributionRadiation Dose EstimationSource Dosimetry
The Sievert integral, widely used to compute dose distributions about filtered line sources, assumes that the emitted energy fluence is exponentially attenuated by the filter thickness traversed by the primary photons. To evaluate this approximation, a Monte Carlo simulation was performed realistically modelling the diffusion and energy degradation of primary photons due to coherent, incoherent and photoelectric interactions in the source filter. Estimates of the exposure rate at points near the source were obtained using analytical averaging. Comparison of the two models shows that for 226Ra and 192Ir sources, the Sievert algorithm consistently overestimates the exposure rate per unit activity. However, such errors may be significantly reduced if source intensity is expressed in terms of exposure rate. Computed exposure rate distributions based on exposure rate calibrations are also less sensitive to uncertainties in available spectroscopic data.
9
On the Estimation of Flux at a Point by Monte Carlo
M. H. Kalos · Nuclear Science and Engineering · 1963 · 100 citations
Studies of Self-Absorption in Gamma-Ray Sources
Robley D. Evans, Richard O. Evans · Reviews of Modern Physics · 1948 · 64 citations