Concepedia

Publication | Closed Access

Evaluation and quality assessment of glucose concentration measurement in blood by point-of-care testing devices

18

Citations

7

References

2006

Year

Abstract

In conclusion, this study shows an insufficient trueness of glucose measurements by POCT devices in the normo- and hypoglycemic range. To improve quality assessment, sample splitting and simultaneous measurement of blood glucose concentration every 4 weeks by POCT devices and of plasma glucose concentration by a reliable TMP is recommended.

References

YearCitations