IEEE Transactions on Circuits & Systems II Express Briefs · 2015 · 41 citations · 23 references
EngineeringBioengineeringBiomedical EngineeringBiosensorsElectroanalytical MethodsSensorsBiomedical InstrumentationBiomedical SensorsElectrochemical BiosensorsElectrochemical SensorsImplantable Biomedical DevicesImplantable BiosensorsGlucose BiosensingGlucose BiosensorsGlucose ConcentrationGlucose SensingGlucose SensorsImplantable Glucose SensorsOperational Amplifier CircuitsOperational Amplifier Design
In this brief, a 1-V potentiostat circuit for a glucose sensor has been proposed. Test results of this 1-V potentiostat demonstrate high linearity with respect to glucose concentration, which is comparable with conventional potentiostat designs. This high linearity is achieved with the help of a transconductance enhanced bulk-driven operational amplifier. The overall current consumption of the circuit is around 22 μA. The potentiostat is tested as a glucose biosensor for glucose concentration ranging from 0 to 30 mM. The design is realized in a 0.35-μm bulk-CMOS process, and the area for the core layout is 0.13 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> . The proposed low-power potentiostat can be incorporated in an implantable glucose sensor system due to its low-power operation and small footprint.
23