Thorax · 1978 · 25 citations · 17 references
MedicineCardiologySurgeryCardiothoracic SurgeryUltrasonographyUltrasoundCardiopulmonary BypassExtracorporeal CirculationHemodynamicsAir BubblesArterial Line PlacementBlood Filtration PropertiesComparative Efficiency AnalysisEfficiencyFiltersMicrocirculationUltrasonic Detection
The ability of 12 commercially available cardiotomy reservoirs to remove bubbles from aspirated blood was investigated by means of a simulated cardiopulmonary bypass circuit and an ultrasonic microbubble detector. Performance varied considerably. The number of gaseous microemboli remaining after passage of blood through the reservoir was reduced by (a) holding the blood in the reservoir, (b) reducing the volume of air mixed with the aspirated blood, and (c) using a reservoir that did not induce turbulence and that contained integral micropore filtration material. Further micropore filtration of the blood after passage through the cardiotomy reservoir was beneficial, and significantly more bubbles were extracted when the microfilter was sited below the reservoir than when it was placed in the arterial line.
17
ACRODERMATITIS ENTEROPATHICA: A LETHAL INHERITED HUMAN ZINC-DEFICIENCY DISORDER
E J Moynahan · The Lancet · 1974 · 627 citations
Medicine, Dermatology, Pathology +7
Central Nervous System Dysfunction Following Open-Heart Surgery
Henry M. Tufo · JAMA · 1970 · 290 citations
Prevention of neurological damage during open-heart surgery.
M A Branthwaite · Thorax · 1975 · 161 citations · Full text
Cerebral protection during open-heart surgery.
T Åberg, Margareta Kihlgren · Thorax · 1977 · 112 citations · Full text