Neck fracture of a cementless forged titanium alloy femoral stem following total hip arthroplasty: a case report and review of the literature

Theodoros B Grivas, Olga D. Savvidou, Spyridon A Psarakis, Pierre-Francois Bernard, George Triantafyllopoulos, Ioannis Kovanis, Panagiotis Alexandropoulos

Journal of Medical Case Reports · 2007 · 33 citations · 16 references

DOIFull text

Open access

Concepts

Abstract

INTRODUCTION: Fractures of the neck of the femoral component have been reported in uncemented total hip replacements, however, to our knowledge, no fractures of the neck of a cementless forged titanium alloy femoral stem coated in the proximal third with hydroxy-apatite have been reported in the medical literature. CASE PRESENTATION: This case report describes a fracture of the neck of a cementless forged titanium alloy stem coated in the proximal third with hydroxy-apatite. CONCLUSION: The neck of the femoral stem failed from fatigue probably because of a combination of factors described analytically below.