The Effect of Bipole Tip-to-Ring Distance in Atrial Electrodes upon Atrial Tachyarrhythmia Sensing Capability in Modern Dual-Chamber Pacemakers
11
Citations
30
References
2009
Year
Atrial electrodes with a short TTR (<10 mm) significantly reduce FFRWO without increasing undersensing and should be used routinely in patients with paroxysmal atrial tachyarrhythmias. However, 20% of atrial tachyarrythmia episodes were incorrectly classified as terminated by these modern devices due to undersensing. Clinicians should be wary of using device-derived endpoints that rely on AF episode number or duration as these may be falsely increased or reduced, respectively.