X-Ray Spectrometry · 1991 · 18 citations · 5 references
Natural SciencesChemistryFood ScienceAnalytical ChemistryElemental AnalysisFood AnalysisTrace AnalysisTrace Element AnalysisTrace Element ChemistryX-ray DetectionX-ray Fluorescence AnalysisX-ray SpectroscopyAnalytical Detection LimitsEnergy Dispersive X-ray SpectroscopyEnergy-dispersive X-ray SpectroscopyMeasurement System DesignMeasurement System ImplementationReference MaterialsStandard Reference MaterialsX-ray Fluorescence Detection
Abstract A rapid method for the simultaneous determination of trace elements in milk powder by energy‐dispersive x‐ray fluorescence has been developed. The measurement system consists of an Si(Li) detector, an x‐ray tube coupled with zirconium as a secondary target excited source and a personal computer. Milk powder was analysed as pressed pellets and irradiated for 1000 s. Concentrations were determined by a fundamental parameters method. The detection limits are close to 1 ppm for most of the trace elements. The accuracy of the method was ascertained by analysing standard reference materials from the International Atomic Energy Agency. Good agreement was found between the certified and experimental values. Potential areas of applicability of the method are indicated.
5
Trace Element Analysis in Thick Organic Specimens by Photon Excited X-ray Fluorescence
G. Alenius, H. Riedl, A. Rindby et al. · Physica Scripta · 1977 · 11 citations
Calibration of the Radioisotope-Excited X-Ray Spectrometer with Thick Standards
C. T. Yap, Peter Kump, S.M. Tang et al. · Applied Spectroscopy · 1987 · 10 citations