A method for monitoring and controlling reproducibility of intensity data in complex electrospray mass spectra: A thermometer ion-based strategy
19
Citations
47
References
2008
Year
Thermometer IonsSpectrochemical AnalysisThermometer Ion-based StrategyAnalytical InstrumentationBioanalysisBiostatisticsAnalytical ChemistryClinical ChemistryInstrumentationMicroarray Data AnalysisBiophysicsBiochemistryIntensity DataIon MobilityNatural SciencesSpectroscopyMass Spectral DataMass SpectrometryInstrument PerformanceMedicine
Reproducibility in mass spectral data is important in both biomarker discovery and spectral database searching. We report a strategy, employing a series of substituted benzylpyridinium thermometer ions that can be used to monitor changes in performance of multiple aspects of an electrospray ionization source that impact the intensity axis of a spectrum. Performance attributes, which could confound even isotope-based quantification strategies, are readily assessed using a mixture of thermometer ions. Based on the observed behavior of the ions, a procedure is proposed for monitoring instrument performance and compensating for factors that affect reproducibility across both time and instruments.