EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsDiagnostic ImagingBiomedical InstrumentationLight-tissue InteractionMedical Imaging InstrumentationMedical InstrumentationNear-infrared SpectroscopyOptical SpectroscopyFunctional Near-infrared SpectroscopyPhoton MigrationSignal-to-noise RatioSignal-to-noise Ratio AnalysisSignal-to-noise Ratio EnhancementSolid PhantomsTime-domain Near-infrared SpectroscopyWavelength Multiplexing
We have designed a compact dual wavelength (687 nm, 826 nm) multi-channel (16 sources, 8 detectors) medical device for muscle and brain imaging based on time domain functional near infrared spectroscopy. The system employs the wavelength space multiplexing approach to reduce wavelength cross-talk and increase signal-to-noise ratio. System performances have been tested on homogeneous and heterogeneous tissue phantoms following specifically designed protocols for photon migration instruments. Preliminary in vivo measurements have been performed to validate the instrument capability to monitor hemodynamic parameters changes in the arm muscle during arterial occlusion and in the adult head during a motor task experiment.