Medical Physics · 1981 · 22 citations · 0 references
MedicineOncologyRadiation OncologyRadiation DosimetryRadiation MeasurementRadiation PhysicsAbsorbed DoseElectron DosimetryExternal Beam DosimetryPhoton DosimetryRadiation Dose MeasurementRadiation TherapyBeam Delivery SystemsClinical Photon Beam DosimetryEntrance Surface DoseIonizing Radiation DosimetryParticle Beam DosimetryPhoton Dose DistributionPhoton RadiotherapySurface Dose Measurement
Buildup and surface-dose measurements were taken for the 6 MV photon beam from a Therac 6 linear accelerator manufactured by Atomic Energy of Canada Limited (AECL) with and without a lucite blocking tray in place. Further measurements were made with a copper filter designed to reduce secondary electrons emitted by photon interactions with the Lucite tray. The results are discussed in relation to skin-sparing for radiation therapy patients. The measurements were made with a fixed volume PTW parallel-plate ionization chamber and corrected to zero-chamber volume. The results were found to be consistent with similar measurements taken with a variable volume extrapolation chamber.