Inappropriate Management of Self‐Terminating Ventricular Arrhythmias by Implantable Cardioverter Defibrillators Despite a Specific Reconfirmation Algorithm: A Report of Two Cases

TANJA M. HURST, HOLGER KRIEGLSTEIN, Harald Tillmanns, Bernd Waldecker

Pacing and Clinical Electrophysiology · 1997 · 11 citations · 2 references

Concepts

Abstract

Algorithms that attempt to reconfirm the presence of an arrhythmia prior to definite treatment have been implemented in ICDs to prevent inappropriate shock therapy due to self-terminating ventricular arrhythmias. Nevertheless, in two patients, clinically inappropriate shocks were delivered after spontaneous conversion of the arrhythmia despite the use of a specific reconfirmation algorithm. Reconfirmation criteria were met due to a premature ventricular complex causing a short cycle in the first patient and a long postextrasystolic pause in the second patient. To avoid inappropriate shock therapy due to self-terminating ventricular arrhythmias, further improvement of detection algorithms is required.