Physics in Medicine and Biology · 1971 · 23 citations · 7 references
MedicineNuclear EngineeringRadiation DosimetryRadiation MeasurementRadiation PhysicsAbsorbed DoseDosimetry InstrumentationIonization Chamber DosimetryIonization ChambersRadiation Dose MeasurementExposure MeasurementGamma DosimetryIonization Chamber ConstructionIonization Chamber DesignIonization Chamber PhysicsIonization Chamber TheoryIonizing Radiation DosimetryNeutron Equivalent Dose
A cylindrical ionization chamber with polythene walls, central electrode, and thin carbon conducting surfaces is described. Its calibration for measuring absorbed dose in a medium irradiated with a fast neutron beam is discussed. The dose to soft tissue is given by DT=0.79xc/RcRT, where Rc is the reading obtained with the chamber given an exposure xc roentgens of 60Co gamma -rays, and RT is the reading obtained in the neutron beam. It is estimated that the uncertainty on DT is +or-6%.
7
Energy Transfer to Matter by Neutrons
R. L. Bach, Randall S. Caswell · Radiation Research · 1968 · 99 citations
An Absolute Tissue Dosemeter for Fast Neutrons
G. S. Hurst · British Journal of Radiology · 1954 · 35 citations