Publication | Closed Access
Biomechanical Evaluation of the Dynamic Hip Screw With Two- and Four-Hole Side Plates
96
Citations
17
References
2000
Year
Peak load in the failure test was not found to be statistically different between the two-hole and four-hole designs. In cyclic testing, the two-hole configuration exhibited statistically smaller fragment migration in both shear and distraction than the four-hole design (p < 0.05). The strain magnitude in the side plate was not statistically different in the cyclic or failure tests. The femurs with a greater neck angle failed by crushing of the bone in the neck. The femurs with a lesser neck angle failed due to bending of the hardware. The results of this investigation revealed that the two-hole DHS is biomechanically as stable as the four-hole DHS in cyclic and failure loads under the conditions tested. These results, in concert with clinical experience, can be used to support the use of the two-hole DHS for the reconstruction of intertrochanteric fractures without a diaphyseal extension.
| Year | Citations | |
|---|---|---|
Page 1
Page 1