Journal of Refractive Surgery · 2011 · 10 citations · 18 references
MedicineOphthalmologySurgeryMicrosurgeryOphthalmic SurgeryReconstructive SurgeryFlap SalvageLaser SurgeryPostoperative Refractive OutcomesPreoperative Refractive EvaluationRetinal SurgeryFemtosecond Laser PulsesFemtosecond Laser UseFemtosecond Laser-assisted KeratoplastyFlap CreationLaser Surgery ComplicationsPostoperative Visual AcuityPostoperative Visual LossUncorrected Visual Acuity
PURPOSE: To evaluate risk factors, management, and outcomes in patients who experienced suction loss during LASIK flap creation with a femtosecond laser (Intra-Lase FS 60, Abbott Medical Optics Inc). METHODS: Pre- and postoperative visual outcomes and refractions for 71 eyes of 70 patients after suction loss during flap creation with the IntraLase FS60 femtosecond laser were analyzed and studied. RESULTS: In patients requiring a second or third lamellar flap cut immediately after suction loss, 97.2% of eyes achieved uncorrected distance visual acuity of 0.00 logMAR (20/20) or better and 100% of eyes achieved corrected distance visual acuity of 0.00 logMAR (20/20) or better at most recent follow-up. CONCLUSIONS: Our study suggests that excellent visual outcomes can be achieved if an immediate lamellar recut is performed after suction loss with a femtosecond laser.
18
Five-Year Outcome of LASIK for Myopia
Naoko Kato, Ikuko Toda, Yoshiko Hori-Komai et al. · Ophthalmology · 2007 · 96 citations
Medicine, Ophthalmology, Surgery +12