Speciation analysis of chromium in environmental water samples from Niigata Prefecture by using ion-exchange/ICP-AES

Yu‐ki Iwasaki, Naoki Kano, Omori Kazuhiro, Hiroshi Imaizumi, Toshio Ishizuka

BUNSEKI KAGAKU · 2004 · 12 citations · 7 references

DOIFull text

Open access

Concepts

Abstract

A simple and an efficient preconcentration method for the separate determination of each species of chromium {i.e., Cr(III) and Cr(VI)} in environmental water samples has been investigated, followed by inductively coupled plasma-atomic emission spectrometry (ICP-AES). This method is based on the following ion-exchange method. Cr(III) was adsorbed on cation-exchange resin, and Cr(VI) on anion-exchange resin. The retained Cr(III) and Cr(VI) were quantitatively eluted with 2 mol dm−3 hydrochloric acid (HCl) and 2 mol dm−3 potassium chloride (KCl), respectively. The preconcentration factor was 100 for Cr(III) and 300 for Cr(VI) in this work, and the detection limits were 0.03 μg dm−3 in Cr(III) and 0.01 μg dm−3 in Cr(VI). The method has been applied to environmental water samples from Niigata Prefecture. The recoveries of Cr(III) and Cr(VI) in environmental water samples were all 94∼98%, and the effect of interference from the coexistence ion was not detected. Therefore, the above-mentioned method is applicable to environmental water samples.

References

7