Concepedia

Publication | Open Access

A Prospective Evaluation of Two Defibrillation Safety Margin Techniques in Patients with Low Defibrillation Energy Requirements

26

Citations

20

References

1998

Year

Abstract

This study is the first to compare prospectively the defibrillation efficacy of two defibrillation safety margins. In patients with a defibrillation energy requirement < or = 6 J, a higher rate of successful defibrillation is achieved with a safety margin of 10 J than with a safety margin equal to the defibrillation energy requirement.