Intramedullary nailing of femoral shaft fractures. Part I: Decision-making errors with interlocking fixation.

Robert J. Brumback, John P. Reilly, A Poka, R Lakatos, G H Bathon, Andrew R. Burgess

Orthopedic Trauma Directions · 2005 · 97 citations · 0 references

Concepts

Abstract

Author summary Of 133 dynamic femoral intramedullary nailing (IM) procedures performed after locking techniques became available: 10.5% (n = 14) were complicated by loss of postoperative fixation and reduction. 13 of the 14 femora shortened an average of 2.0 cm and one shortened slightly with clinical loss of rotational stability. Errors in surgical judgment were attributed to (alone or in combination): Inadequate preoperative analysis of fracture patterns Undetected comminution during reaming or nail insertion Failure to recognize postoperatively increased comminution Fracture instability Authors suggest: Use of high-quality preoperative x-rays to detect nondisplaced comminution of major fracture fragments. Intraoperative fluoroscopic assessment: increase in comminution secondary to reaming or nail insertion is an indication for static locked fixation. Analysis of radiographs taken immediately postoperatively while patient is still under anesthesia to evaluate previously undetected instability that should be treated by static locked fixation. Dynamic IM stabilization should be reserved for transverse or short oblique fractures at the femoral isthmus that have type I or type II comminution.