Publication | Open Access
Dose Validation of Physical Wedged Asymmetric Fields in Artiste Linear Accelerator
15
Citations
20
References
2014
Year
EngineeringLinear AcceleratorSuperposition AlgorithmSurgeryAccelerator PhysicBrachytherapyArtiste Linear AcceleratorRadiation Therapy PlanningInstrumentationRadiation OncologyNuclear MedicineRadiologyAccelerator TechnologyPhysicsRadiation TherapyMedical ImagingMedicineAbsorbed DoseSynchrotron RadiationDosimetryRadiation DoseOncologyDose ValidationParticle AcceleratorBeam Transport System
Aim: The purpose of this study was to make a comparison between measured and calculated physical wedge dose distributions using the superposition algorithm. Settings and Design: The accurate determination of absorbed dose is important radiotherapy because of the relatively steep sigmoidal dose response curves for both tumor control and normal-tissue damage. Materials and Methods: High-energy photons (6 and 10 MV) from Artiste Treatment System Linear Accelerator Machine, available at Alexandria Ayadi Al-Mostakbal Oncology Center, were used. Results and Discussion: The results showed that the difference between measured and calculated wedged isodose curves depends on field size, beam energy, and the angle of the used wedge. Conclusion: The results showed that the presence of a wedge alters the primary and scattered components generated by a linear accelerator and causes beam hardening in 6 and 10 MV. The beam hardening increased as the wedge angle increased.