Circulation Journal · 2008 · 69 citations · 4 references
EngineeringCardiologyBiomedical ImagingBiomedical OpticsDiagnostic ImagingCoronary Artery ImagingImage Quality AssessmentOptical CoherenceOptical Coherence TomographyOptical TomographyImage Acquisition Quality
BACKGROUND: Because optical coherence tomography (OCT) requires balloon-occlusion to overcome image attenuation by blood, the present study investigated the safety and usefulness of a new non-occlusive technique. METHODS AND RESULTS: The 40 angina patients were assigned to the continuous-flushing method or the balloon-occlusion method group. The continuous-flushing method was superior for observing proximal lesions compared with the balloon-occlusion method (94% vs 55%, p=0.01). There were no differences between groups in the visible length and image quality. There were no major complications during OCT. CONCLUSIONS: The continuous-flushing method is safe and useful for OCT.
4
Ik‐Kyung Jang, Brett E. Bouma, Dong-Heon Kang et al. · Journal of the American College of Cardiology · 2002 · 1K citations
From Bench to Bedside A Novel Technique of Acquiring OCT Images
Francesco Prati, Maria Cera, Vito Ramazzotti et al. · Circulation Journal · 2008 · 263 citations · Full text
Jun Tanigawa, Peter Barlis, Carlo Di Mario · Circulation Journal · 2007 · 73 citations · Full text
Medicine, Cardiology, Surgery +17