A Fast Coincidence-Anticoincidence Analyzer

Richard L. Garwin

Review of Scientific Instruments · 1953 · 56 citations · 2 references

Concepts

Abstract

A modified Rossi circuit has been extended to a form in which it allows the recording of as many as six coincidence and four anticoincidence pulses with a resolving time of 3×10−9 second. All input pulses are negative and of any amplitude > 3 volts. A complete schematic is given of the circuit which has operated for six months without possibility of adjustment.

References

2