Pacing and Clinical Electrophysiology · 1983 · 44 citations · 7 references
MedicineCardiologyElectrical EngineeringBioinstrumentationElectrophysiologyInterventional CardiologyAtrial ElectrophysiologyAtrial FibrillationCardiac ArrhythmiasCardiac PacingElectrode PlacementIntracardiac ElectrophysiologyAtrial Electrogram AnalysisIntracardiac PacingPacing In Atrial FibrillationSlew RateSurface Electrocardiography
In order to explore fully the sensing characteristics of both bipolar and unipolar configurations of the tined right atrial appendage electrode, detailed electrographic analysis was performed in 10 patients undergoing atrial electrode implantation. Though no differences were noted between unipolar and bipolar atrial deflections in electrograms from these leads, the unipolar ventricular deflections were greater in both amplitude (1.9 vs 0.6 mV) and slew rate (0.17 vs 0.07). These data suggest that bipolar pacing configurations might be preferable when selective atrial sensing is important.
7
Clinical Experience with a New Transvenous Atrial Lead
Paul Citron, Nicholas P.D. Smyth, Manfred Kleinert et al. · CHEST Journal · 1978 · 39 citations
Medicine, Cardiology, Surgery +5
Clinical experience with transvenous atrial pacing.
J S Geddes, S W Webb, Ian P. Clements · Heart · 1978 · 28 citations · Full text