Concepedia

Publication | Closed Access

Motion as a perturbation: Measurement-guided dose estimates to moving patient voxels during modulated arc deliveries

21

Citations

25

References

2013

Year

Abstract

The dose to an arbitrary moving voxel in a patient can be estimated with acceptable accuracy for a VMAT delivery, by performing a single QA measurement with a cylindrical phantom and applying two consecutive perturbations to the TPS-calculated patient dose. The first one accounts for the differences between the planned and delivered static doses, while the second one corrects for the motion.