Optics Express · 2005 · 80 citations · 15 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsDiagnostic ImagingDoppler UltrasonographyImagingMedical ImagingOptical CoherenceOptical Coherence TomographyOptical TomographyBlood FlowCross-sectional ImagingDoppler Optical Coherence TomographyFrequency EstimationMeasurement Precision EstimationPrecisionPrecision EstimationVascular Flow Imaging
Doppler optical coherence tomography (DOCT) is a technique for simultaneous cross-sectional imaging of tissue structure and blood flow. We derive the fundamental uncertainty limits on frequency estimation precision in DOCT using the Cramer-Rao lower bound in the case of additive (e.g., thermal, shot) noise. Experimental results from a mirror and a scattering phantom are used to verify the theoretical limits. Our results demonstrate that the stochastic nature of frequency noise influences the precision of flow imaging, and that the noise model must be selected judiciously in order to estimate the frequency precision.
15
David Huang, Eric A. Swanson, Charles P. Lin et al. · Science · 1991 · 13.7K citations
Optical Doppler tomographic imaging of fluid flow velocity in highly scattering media
Zhongping Chen, Thomas E. Milner, Digant P. Davé et al. · Optics Letters · 1997 · 441 citations · Full text
Engineering, Physics, Radiology +17