Determination of multislice computed tomography dose index (CTDI) using optically stimulated luminescence technology
47
Citations
16
References
2010
Year
The LED-based OSL system can be an accurate alternative device for CT dose evaluations. CTDI100 measurement with the use of a 100 mm pencil ionization chamber substantially underestimates the CTDIinfinity value even with 5 mm collimated beam width. The established complete set of CTDI(100-efficiency) correction factors for various scan parameters allows for accurately estimating CTDIinfinity with the current use of pencil chamber and dose phantoms. Combined with the simple calibration, it gives this work great potential to be used not only in routine clinical quality assurance checks but also as a promising tool for patient organ dose assessment.