MedicineRadiologyRadiation DosimetryRadiation MeasurementRadiation ProtectionDose CalibrationIonization Chamber DosimetryIonization ChambersPhantom DosimetryRadiation Dose MeasurementRadiation InstrumentationDosimetry Phantom DesignExposure MeasurementIn-air DosimetryInstrument CalibrationPhantom-based Dose MeasurementSolid PhantomsThimble Ionization ChambersX-ray Dosimetry
The thimble ionization chamber-condenser r-meter was designed (10, 11) for the measurement of the radiation quantity now known as exposure dose. It was intended for use in air with a broad, substantially parallel beam of x-rays and is calibrated with such a beam. Over the years, however, this instrument has also been exposed, while immersed in tissue-equivalent phantoms and other condensed media, to elicit depth-dose information, among other things. Apparently the assumption involved is that the thimble will record faithfully (say, within 5 per cent) the exposure dose which would have been delivered at the position of the center of the chamber but in its absence. It is the purpose of this paper to examine the validity of this assumption for the Victoreen condenser r-meter over a wide range of quality and other physical factors. It might be well to lead up to the discussion of corrections applicable to readings obtained in phantoms by reviewing corrections which may be applicable to readings resulting from exposures in air. Atmospheric Conditions Since the thimble is not hermetically sealed, the number of molecules in the active volume will fluctuate with the temperature and pressure of the atmosphere. The instrument calibration is referred to an atmospheric temperature of 22° C. and pressure of 760 mm. Hg. If a reading R is obtained at (T, P), the corrected reading would be. Relative humidity has been mentioned as a source of error.2 The slight correction applicable for water vapor in the active volume is discussed in Section 2.1 of NBS Handbook 64 (30). Surface conductivity (7) may be greatly increased on the insulators as a result of high relative humidity and, in fact, this effect is used in one class of relative humidity transducers (19). No correction has been determined for this effect; rather, the problem is usually avoided by storing the instrument in a dry atmosphere. Thimble-Electrometer Details Part of the art of manufacturing a successful thimble chamber involves the inclusion of certain “impurities” on the electrodes in the active volume. This enables the response relative to that of a standard free-air chamber to be made nearly constant over a useful range of x-ray qualities. There remains some variation over the recommended quality range. Relatively large variations exist outside this range (owing to wall absorption on the low-energy side and to lack of electron build-up on the high-energy side). If the thimble is not used with the electrometer with which it was calibrated, or if the electrometer characteristics have changed since calibration, additional corrections will be applicable. It is customary to accumulate all such corrections into a single factor, F, which is multiplicative with the above Rc to express the exposure in roentgens.
5
Spectral Distributions of 280 kVp X Rays
D. V. Cormack, W. E. Davitt, D.G. Burke et al. · British Journal of Radiology · 1958 · 21 citations