Medical Physics · 1979 · 41 citations · 0 references
Natural SciencesNuclear EngineeringPhysicsNuclear PhysicsRadiation MeasurementRadiation PhysicsHeavy-ion PhysicsNeutron DetectionNeutron Spectrum AnalysisNuclear SpectroscopyNuclear StructureRadiation SpectroscopyEnergy Spectra MeasurementNeutron Emission SpectraNeutron Energy SpectraNeutron Energy SpectrumNeutron Equivalent DoseNeutron SpectrometryNeutron SpectroscopyNeutron Spectrum Characterization
The energy spectra of neutrons produced by 35 and 46 MeV protons, 16 and 28 MeV deuterons, and 44 MeV 3He ions on thick beryllium were measured at angles of 0 degrees, 15 degrees, and 45 degrees with respect to the incident beams. The spectra were measured by the time-of-flight method for neutrons from the maximum energy down to 1 MeV. Neutron dose rates obtained from the zero-degree spectra by use of available tissue kerma factors agree with TE-TE ionization chamber measurements.