Scandinavian Journal of Urology and Nephrology · 1976 · 94 citations · 6 references
MedicineSurgeryUrologyFemale UrologyUltrasonographyUrodynamicsBladder AssessmentBladder SurgeryUrethral FunctionUrinary Continence AssessmentUrinary IncontinenceUrodynamic EvaluationBladder Pressure MeasurementBladder Pressure RecordingIntravesical Pressure DynamicsIntravesical Pressure MeasurementMaximum Urethral Closure PressureUrodynamic Pressure Measurement
A new technique for studying simultaneous intravesical-intraurethral pressure is described. The basic principle of our technique is pressure recordings with the aid of micro-transducers. These micro-transducers have excellent frequency response, are easy to calibrate and to sterilize and are free from artefacts induced by motion. They permit a more accurate study of the pressures within the bladder and the urethra than catheter-transducer systems described earlier. Using electronical subtraction the closure pressure can easily be measured under static and dynamic conditions. Typical pressure diagrammes obtained with this catheter system are presented.
6
A catheter-manometer calibrator--a new clinical instrument.
M. Asmussen, K. Lindström, U. Ulmsten · PubMed · 1975 · 23 citations