Investigation of radiological properties and water equivalency of PRESAGE<sup>®</sup> dosimeters
86
Citations
40
References
2011
Year
Based on the results of this study, the new PRESAGE formulations with lower halogen content are more radiologically water equivalent overall than the original formulation. This indicates that the new PRESAGE formulations are better suited to clinical applications and are more accurate dosimeters and phantoms than the original PRESAGE formulation. While correction factors are still needed to convert the dose measured by the dosimeter to an absorbed dose in water in the kilovoltage energy range, these correction factors are considerably smaller for the new PRESAGE formulations compared to the original PRESAGE and the existing polymer gel dosimeters.