Physics of Particles and Nuclei Letters · 2014 · 30 citations · 9 references
EngineeringNuclear EngineeringNuclear InstrumentationNuclear PhysicsRadiation MeasurementActivation AnalysisActivity MeasurementGamma-ray InstrumentationGamma-ray SpectroscopyNeutron Activation AnalysisReactor InstrumentationActivation Energy AnalysisInstrumental Neutron Activation AnalysisNeutron Detection PrinciplesSystem Automation
The automation system for measurement of induced activity of gamma-ray spectra for multi-element high volume neutron activation analysis (NAA) was designed, developed and implemented at the reactor IBR-2 at the Frank Laboratory of Neutron Physics. The system consists of three devices of automatic sample changers for three Canberra HPGe detector-based gamma spectrometry systems. Each sample changer consists of two-axis of linear positioning module M202A by DriveSet company and disk with 45 slots for containers with samples. Control of automatic sample changer is performed by the Xemo S360U controller by Systec company. Positioning accuracy can reach 0.1 mm. Special software performs automatic changing of samples and measurement of gamma spectra at constant interaction with the NAA database.
9
P. Bode, Robert R. Greenberg, Elisabete A. De Nadai Fernandes · CHIMIA International Journal for Chemistry · 2009 · 48 citations · Full text
A. Yu. Dmitriev · Physics of Particles and Nuclei Letters · 2013 · 40 citations