Natural SciencesChemical EngineeringChemistryAnalytical ChemistryElectroanalytical MethodsElectrochemistryAquatic ChemistryElectroanalytical ChemistryElectrochemical DetectionElectrochemical SensorsEnvironmental Analytical ChemistryStripping AnalysisAdsorptive Stripping VoltammetryElectrochemical MeasurementsElectrode Ion AdsorptionHeavy Metal DetectionMethod Detection LimitPreconcentrationSample PreconcentrationSeawater Chemistry
Abstract A highly sensitive electrochemical stripping procedure for uranium is reported. Accumulation is achieved by controlled adsorption of the uranium/4‐(2‐pyridylazo) resorcinol complex on the static mercury drop electrode. The effect of various operational parameters on the stripping response is discussed. The detection limit is 0.2 nM with 10 min preconcentration. The response is linear over a 9.3 × 10 −9 −3.0 × 10 −6 M concentration range, and the relative standard deviation (at the 24 μg·l −1 level) is 3%. Most cations and anions do not interfere in the determination of uranium; some interferences are eliminated by the addition of CDTA and EDTA. The applicability of the method to fresh water and seawater is demonstrated (the results obtained for uranium in reference seawater samples, NASS−2, were 2.95 μg·l −1 with standard deviations of 0.20 μg·l −1 (n = 5).
8