Medical Physics · 1994 · 10 citations · 0 references
MedicineBiomedical EngineeringRadiation DosimetryRadiation MeasurementSemiconductor DevicesAbsorbed DoseMonte Carlo DosimetryPhantom DosimetryRadiation Dose MeasurementRelative DosimetryThermoluminescent DosimetryMonte Carlo Phantom DosimetryPhantom-based Dose MeasurementSemiconductor Detector DosimetrySilicon DosimetryThermoluminescent DetectorsThermoluminescent DosimetersThermoluminescent MaterialsWater Phantom Measurements
The relative dose distribution around the 103Pd model 200 implant seed was measured with a computerized data acquisition system employing a p-n junction silicon diode immersed in a water phantom. Data are acquired in polar coordinates by computer control of (1) the diode distance from the seed center and (2) the rotation angle of seed about a transverse axis. Transverse axis data are compared with thermoluminescent dosimeter (TLD) measurements and a Monte Carlo calculation by others.