Journal of Cataract & Refractive Surgery · 1996 · 31 citations · 9 references
MedicineOphthalmologyOptometryClinical OptometryClinical Vision ScienceOphthalmic SurgeryOcular BiometryRefractive Error AssessmentInterlaboratory ComparisonsInterlaboratory ReproducibilityInterlaboratory StudyLens Power DeterminationRepeatabilityRepeatability AssessmentStandardsTolerances
PURPOSE: To describe a testing program conducted by a standards group as a guide for setting international tolerances for intraocular lens (IOL) dioptric power. SETTING: Multicenter study. METHODS: Seven biconvex, poly(methyl methacrylate) IOLs ranging in power from 10.00 through 30.00 diopter (D) were circulated among nine participating laboratories experienced in IOL optical measurements. Each laboratory performed repeated optical tests to determine dioptric power. These results were analyzed for repeatability and reproducibility in accordance with methods specified by the International Organization for Standardization. RESULTS: Intralaboratory repeatability was less than 0.5% of the dioptric power, and interlaboratory reproducibility, when following a normalized procedure for correction and conversion, was less than 1.0% of the dioptric power. CONCLUSION: Tolerance limits of +/0 0.30 D in the range 0 to 15.00 D, +/- 0.40 D for 15.50 to 25.00 D, and +/- 0.50 D for 25.50 to 30.00 D have been proposed as an international standard for IOLs. The contribution of IOL power error within the limits of the standard are estimated to contribute less than 1.0% to the total error in postoperative refractive prediction.
9
Bradford Cannon, Edited by Stewart Duke-Elder · Plastic & Reconstructive Surgery · 1973 · 2K citations
Sources of error in intraocular lens power calculation
Thomas Olsen · Journal of Cataract & Refractive Surgery · 1992 · 496 citations
Medicine, Ophthalmology, Surgery +17
Elements of Modern Optical Design
John Macdonald · Optica Acta International Journal of Optics · 1986 · 163 citations