The Occupational Therapy Journal of Research · 1987 · 37 citations · 10 references
MedicineKinesiologyRehabilitation MedicineExercise TestingPhysical TherapyPsychometricsExercise TherapyMuscle Strength AssessmentMuscle Strength TestingMuscular Strength TestingPhysical Performance AssessmentPhysical RehabilitationHand-held DynamometryHandgrip Strength TestingNormative Data CollectionNormative Data EstablishmentNormative Health DataNormative Strength DataTreatment Effectiveness AnalysisTreatment Effectiveness Measurement
These two studies evaluated whether three types of Jamar dynamometers measured equivalently. In study A, 26 subjects were tested on 2 types; in study B, 30 subjects were tested on 3 types. Both studies found significant differences between and among the dynamometers. Therapists can no longer assume that all versions of the Jamar dynamometer measure equivalently, and the differences should be taken into account when using normative data and when assessing treatment effectiveness. In the latter case, the same dynamometer must be used for pre- and posttesting.
10
Grip and pinch strength: normative data for adults.
Virgil Mathiowetz, Nancy Kashman, Gloria Volland et al. · PubMed · 1985 · 2.2K citations
Reliability and validity of grip and pinch strength evaluations
Virgil Mathiowetz, Karen C. Weber, Gloria Volland et al. · The Journal Of Hand Surgery · 1984 · 1.9K citations
Clinical assessment Recommendations
Elaine Ewing Fess · Medical Entomology and Zoology · 1981 · 1.2K citations
M Kellor, Jo Frost, Norman E. Silberberg et al. · PubMed · 1971 · 401 citations
Engineering, Anatomy, Grasping +11
Grip strength as measured by the Jamar dynamometer.
Schmidt Rt, Toews Jv · PubMed · 1970 · 325 citations