Temperature response of biological materials to pulsed non‐ablative CO<sub>2</sub> laser irradiation

Marco J. P. Brugmans, J.S. Kemper, Geert H. M. Gijsbers, Freerk W. van der Meulen, Martin J. C. van Gemert

Lasers in Surgery and Medicine · 1991 · 67 citations · 14 references

Concepts

Abstract

This paper presents surface temperature responses of various tissue phantoms and in vitro and in vivo biological materials in air to non-ablative pulsed CO2 laser irradiation, measured with a thermocamera. We studied cooling off behavior of the materials after a laser pulse, to come to an understanding of heat accumulation and related thermal damage during (super) pulsed CO2 laser irradiation. The experiments show a very slow decay of temperatures in the longer time regime. This behavior is well predicted by a simple model for one-dimensional heat flow that considers the CO2 laser radiation as producing a heat flux on the material surface. The critical pulse repetition frequency for which temperature accumulation is sufficiently low is estimated at about 5 Hz. Although we have not investigated the ablative situation, our results suggest that very low pulse frequencies in microsurgical procedures may be recommended.

References

14