Lasers in Surgery and Medicine · 1991 · 67 citations · 14 references
EngineeringBiomedical EngineeringBiophysicsBiomedical OpticsHeat TransferMedical PhysicsBiomedical Laser ApplicationsLaser CoolingLaser-tissue InteractionPhotothermal TherapyTemperature ManagementTissue OpticsHeat Flux MeasurementHeat Stress AccumulationSolid PhantomsSurface TemperatureTemperature Response ModelingThermal AblationThermal Damage
This paper presents surface temperature responses of various tissue phantoms and in vitro and in vivo biological materials in air to non-ablative pulsed CO2 laser irradiation, measured with a thermocamera. We studied cooling off behavior of the materials after a laser pulse, to come to an understanding of heat accumulation and related thermal damage during (super) pulsed CO2 laser irradiation. The experiments show a very slow decay of temperatures in the longer time regime. This behavior is well predicted by a simple model for one-dimensional heat flow that considers the CO2 laser radiation as producing a heat flux on the material surface. The critical pulse repetition frequency for which temperature accumulation is sufficiently low is estimated at about 5 Hz. Although we have not investigated the ablative situation, our results suggest that very low pulse frequencies in microsurgical procedures may be recommended.
14
C. J. Tranter · Journal of the Mechanics and Physics of Solids · 1959 · 18.5K citations
Optical Constants of Water in the 200-nm to 200-μm Wavelength Region
George M. Hale, Marvin R. Querry · Applied Optics · 1973 · 4.4K citations