Photonic Sensors · 2014 · 18 citations · 8 references
Natural SciencesGeophysicsInstrumentationOpticsSeismologyFiber OpticsOptical FibersOptical InstrumentationOptical InstrumentsOptical SensorsSeismic InstrumentationAcceleration SensorsBandwidthCarrier Phase MeasurementElectro-mechanical CouplingFiber SensorsSeismic SensorsSeismometer CalibrationSensitivity
An all-metal 3-component optical fiber seismometer was proposed and experimentally demonstrated. The theoretical analysis was given based on the electro-mechanical theory. Calibration results showed that the axis sensitivity was about 41 dB (re: 0 dB=1 rad/g) with a fluctuation ±2 dB in the frequency bandwidth of 5 Hz–400 Hz. A transverse sensitivity of about −40 dB was achieved. The fluctuation of the acceleration sensitivity for the three accelerometers in the seismometer was within ±2.5 dB. The minimum phase demodulation detection accuracy of the phase-generated carrier (PGC) was 10−5 rad/√Hz, and the minimum detectable acceleration was calculated to be 90 ng/√Hz.
8
A 3-component fiber-optic accelerometer for well logging
Nan Zeng, C.Z. Shi, Ming Zhang et al. · Optics Communications · 2004 · 67 citations
Engineering, Electrical Engineering, Petroleum Engineering +8