A simple atmospheric pressure room-temperature air plasma needle device for biomedical applications

Xinpei Lu, Zilan Xiong, Fading Zhao, Yubin Xian, Qing Xiong, Weiwei Gong, C. L. Zou, Zhonghe Jiang, Y. Pan

Applied Physics Letters · 2009 · 110 citations · 27 references

Concepts

Abstract

Rather than using noble gas, room air is used as the working gas for an atmospheric pressure room-temperature plasma. The plasma is driven by submicrosecond pulsed directed current voltages. Several current spikes appear periodically for each voltage pulse. The first current spike has a peak value of more than 1.5 A with a pulse width of about 10 ns. Emission spectra show that besides excited OH, O, N2(C–B), and N2+(B–X) emission, excited NO, N2(B–A), H, and even N emission are also observed in the plasma, which indicates that the plasma may be more reactive than that generated by other plasma jet devices. Utilizing the room-temperature plasma, preliminary inactivation experiments show that Enterococcus faecalis can be killed with a treatment time of only several seconds.

References