IEEE Transactions on Nuclear Science · 1995 · 21 citations · 14 references
EngineeringNuclear EngineeringRadiologyDiagnostic ImagingDiagnostic RadiologyRadiation DetectionDetector TechnologyMedical ImagingPositron Emission TomographyRadiation DetectorsRadiation Imaging DetectorsScintillation DetectorsCount Rate PerformanceDead Time SimulationDetector ConstructionDetector DesignPositron Emission Tomography DetectorsRate CapabilityTime ResolutionWhole-body Pet Imaging
The design and construction of the first of two BaF/sub 2/-TMAE large area detectors to be incorporated in a unique double headed whole body PET system (PETRRA) has been completed. Preliminary experimental work carried out with the first large area detector shows an absolute detection efficiency of 20.5% and a time resolution of 3.5 ns (FWHM) and 8.0 ns (FWTM), using 8 mm thick scintillation crystals. An analytical model of the electronic read-out system has been developed to investigate the rate capability of the new scanner. This model predicts that under typical imaging conditions, with 40 MBq of activity in the field of view, the proposed scanner will acquire coincidence data at 20-25 kcps.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
14
Radiation detection and measurement
Nuclear Instruments and Methods · 1980 · 1.6K citations
A rotating PET scanner using BGO block detectors: design, performance and applications.
David W. Townsend, M. Wensveen, L. Byars et al. · PubMed · 1993 · 79 citations
Charles Branch Wilson, Helen Young, R.J. Ott et al. · European Journal of Nuclear Medicine and Molecular Imaging · 1995 · 64 citations
Medicine, Nuclear Medicine, Oncology +14