Optics Express · 2004 · 295 citations · 17 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsDiagnostic ImagingLight-tissue InteractionMedical ImagingOptical Coherence TomographyOptical TomographyScattering In TissueTissue Optical PropertiesAttenuation Coefficient EstimationAttenuation Coefficient MeasurementAttenuation Coefficient ModelsPoint Spread FunctionQuantitative Optical Coherence TomographySingle ScatteringSingle-scattering ModelsTissue Scattering Coefficients
From calibrated, weakly scattering tissue phantoms (2-6 mm-1), we extract the attenuation coefficient with an accuracy of 0.8 mm-1 from OCT data in the clinically relevant 'fixed focus' geometry. The data are analyzed using a single scattering model and a recently developed description of the confocal point spread function (PSF). We verify the validity of the single scattering model by a quantitative comparison with a multiple scattering model, and validate the use of the PSF on the calibrated samples. Implementation of this model for existing OCT systems will be straightforward. Localized quantitative measurement of the attenuation coefficient of different tissues can significantly improve the clinical value of OCT.
17
David Huang, Eric A. Swanson, Charles P. Lin et al. · Science · 1991 · 13.7K citations
Savitzky-Golay Smoothing Filters
William H. Press, Saul A. Teukolsky · Computers in Physics · 1990 · 11.8K citations · Full text
Practical Statistics for Medical Research.
Sophie Walter, Daniel Altman · Biometrics · 1992 · 11.7K citations
Optical coherence microscopy in scattering media
Joseph A. Izatt, Eric A. Swanson, James G. Fujimoto et al. · Optics Letters · 1994 · 628 citations