Energy dependence of thermoluminescent dosemeters for X-ray dose and dose distribution measurements in a mouse phantom
11
Citations
3
References
1974
Year
Mouse-sized Test PhantomDose Distribution MeasurementsRadiation EffectRadiation ExposureRadiation ProtectionRadiation TestingRadiation Therapy PlanningInstrumentationRadiation OncologyNuclear MedicineRadiologyHealth SciencesTl ResponseMedical ImagingDosimetryX-ray DoseDose DistributionRadiation DoseEnergy DependenceMedicine
A mouse-sized test phantom containing TL dosemeters was exposed to X-rays having a HVL value of 1.4-2.8 mm Cu. Dependence of the response of 7LiF dosemeters on the HVL of the incident beam resulted in an increase of TL response per rad in muscle for dosemeters positioned at the central axis of the phantom compared with the response of similar dosemeters exposed free-in-air. An increase of 7.3% in TL response per rad in muscle was found over the HVL range. The spread in effective energy inside the phantom was small and the dose distribution may be estimated directly from the TL response of 7LiF dosemeters at the entrance, central and exit position in the phantom.