Journal of applied meteorology · 1978 · 22 citations · 5 references
Natural SciencesPhysicsAtmospheric ScienceNuclear InstrumentationNuclear PhysicsCloud Condensation NucleiCloud PropertiesCondensationCondensation ProcessesCloud SupersaturationCritical SupersaturationInstrumental Ccn MeasurementLaboratory Cloud Chamber ExperimentsOperating ConditionsSupersaturation
An instrument for measuring the concentrations of cloud condensation nuclei as a function of nucleus critical supersaturation (Se) is described. In the range 0.1% ≤ Se ≤ 1%, the instrument is operated as a vertical flow thermal diffusion chamber. In the range 0.016% ≤ Se ≤ 0.1%, the chamber is operated at 100% humidity, and the equilibrium radii of the nuclei are measured with an optical counter. This paper presents a large body of data taken in the laboratory to evaluate the performance of the instrument. The data indicates that the instrument is internally consistent over a wide range of operating conditions.
5