Analytical Chemistry · 1997 · 41 citations · 32 references
EngineeringChemical EngineeringAnalytical ChemistryBiomemsElectroanalytical MethodsMicrofluidic Device FabricationMicrofluidic DevicesMicrofluidic SystemsMicrofluidicsSurface ImmobilizationCapillary ElectrophoresisCapillary MicrofluidicsInlet DesignMicroextractionSuction Effects
A novel injection scheme is described in which ultrasmall samples in the attoliter (10(-18) L) and low femtoliter (10(-15) L) range, or even single molecules, are controllably introduced into a tapered capillary so that electrophoretic separation can be carried out. To match the dimensions of the capillary inlet with that of the sample, capillary tips are tapered to an inside diameter ranging from hundreds of nanometers to a few micrometers. To inject an ultrasmall sample, optical trapping is used to immobilize and manipulate the sample in order to place it inside or next to the capillary inlet. A small controlled suction results in the loading of the sample into the capillary.
32
Observation of a single-beam gradient force optical trap for dielectric particles
A. Ashkin, J. M. Dziedzic, J. E. Bjorkholm et al. · Optics Letters · 1986 · 7K citations
Optical Trapping and Manipulation of Viruses and Bacteria
A. Ashkin, J. M. Dziedzic · Science · 1987 · 2.2K citations
Nanoscale Magnetic Domains in Mesoscopic Magnets
M. Hehn, Kamel Ounadjela, J. P. Bucher et al. · Science · 1996 · 379 citations