Review of Scientific Instruments · 2014 · 59 citations · 8 references
EngineeringNuclear EngineeringPhysicsNuclear InstrumentationNuclear PhysicsPlasma PhysicsDetector TechnologyNeutron DetectionRadiation Monitoring SystemsReactor InstrumentationLeading-edge DesignNeutron DiagnosticsNeutron FluxNeutron Flux MeasurementNeutron Flux MonitoringNeutron MonitoringPrompt Neutron EmissionResponse Time
A fast time response, wide dynamic range neutron flux monitor has been developed toward the LHD deuterium operation by using leading-edge signal processing technologies providing maximum counting rate up to ∼5 × 10(9) counts/s. Because a maximum total neutron emission rate over 1 × 10(16) n/s is predicted in neutral beam-heated LHD plasmas, fast response and wide dynamic range capabilities of the system are essential. Preliminary tests have demonstrated successful performance as a wide dynamic range monitor along the design.
8
Overview of the Large Helical Device project
A. Iiyoshi, A. Komori, A. Ejiri et al. · Nuclear Fusion · 1999 · 288 citations
Neutron calibration techniques for comparison of tokamak results
J.D. Strachan, J. M. Adams, Cris W. Barnes et al. · Review of Scientific Instruments · 1990 · 36 citations