The Effect of Sensor Malpositioning on Pulse Oximeter Accuracy during Hypoxemia

82

Citations

0

References

1993

Year

Abstract

The calibration curves of the pulse oximeters studied were changed greatly by sensor malpositioning. At low SaO2 values, these changes could cause the oximeter to indicate that a patient was only mildly hypoxemic when, in fact, hypoxemia was profound. It is recommended that sensor position be checked frequently and that inaccessible sensor locations be avoided whenever possible.